FDA Inspection 1006245 — Whitehall/Div of Acorn Engineering Co

Whitehall/Div of Acorn Engineering Co — FEI 1000150265

The FDA completed an inspection of Whitehall/Div of Acorn Engineering Co on March 23, 2017 in City Of Industry, CA. The inspection was classified OAI (Official Action Indicated — objectionable conditions were found and regulatory action is recommended). 16 citations were issued during this inspection. Product area: Devices.

Inspection Details

Classification
Official Action Indicated (OAI) (OAI)
End Date
March 23, 2017
Fiscal Year
2017
Product Type
Devices
Project Area
Compliance: Devices
Location
City Of Industry, CA (United States)

Citations

IDCFRDescription
1471321 CFR 820.198(a)Lack of or inadequate complaint procedures
1472221 CFR 820.40Procedures not adequately established or maintained
235021 CFR 820.25(b)Training - Lack of or inadequate procedures
310221 CFR 820.30(h)Design transfer - Lack of or inadequate procedures
310321 CFR 820.30(i)Design changes - Lack of or Inadequate Procedures
310421 CFR 820.30(j)Design history file
313021 CFR 820.100(a)Lack of or inadequate procedures
315921 CFR 820.184DHR content
333121 CFR 820.181DMR - not or inadequately maintained
334521 CFR 820.200(a)Servicing - Lack of or inadequate procedures
366921 CFR 820.20(c)Management review - defined interval, sufficient frequency
405921 CFR 820.22Quality Audits - defined intervals
54121 CFR 820.70(c)Environmental control Lack of or inadequate procedures
54621 CFR 820.75(a)Lack of or inadequate process validation
63021 CFR 803.17Lack of Written MDR Procedures
73221 CFR 803.50(a)(2)Individual Report of Malfunction