FDA Inspection 1015469 — Alber GmbH

Alber GmbH — FEI 3004730072

The FDA completed an inspection of Alber GmbH on May 11, 2017 in Albstadt. The inspection was classified OAI (Official Action Indicated — objectionable conditions were found and regulatory action is recommended). 4 citations were issued during this inspection. Product area: Devices.

Inspection Details

Classification
Official Action Indicated (OAI) (OAI)
End Date
May 11, 2017
Fiscal Year
2017
Product Type
Devices
Project Area
Compliance: Devices
Location
Albstadt (Germany)

Citations

IDCFRDescription
1471321 CFR 820.198(a)Lack of or inadequate complaint procedures
310121 CFR 820.30(g)Design validation- Lack of or inadequate procedures
310321 CFR 820.30(i)Design changes - Lack of or Inadequate Procedures
313021 CFR 820.100(a)Lack of or inadequate procedures