FDA Inspection 1024822 — Dexcowin Co. Ltd.
The FDA completed an inspection of Dexcowin Co. Ltd. on August 31, 2017 in Geumcheon. The inspection was classified OAI (Official Action Indicated — objectionable conditions were found and regulatory action is recommended). 16 citations were issued during this inspection. Product area: Devices.
Inspection Details
- Classification
- Official Action Indicated (OAI) (OAI)
- End Date
- August 31, 2017
- Fiscal Year
- 2017
- Product Type
- Devices
- Project Area
- Compliance: Devices
- Location
- Geumcheon (Korea (the Republic of))
Citations
| ID | CFR | Description |
|---|---|---|
| 14713 | 21 CFR 820.198(a) | Lack of or inadequate complaint procedures |
| 14716 | 21 CFR 820.30(f) | Design verification - output does not meet input requirement |
| 14722 | 21 CFR 820.40 | Procedures not adequately established or maintained |
| 2327 | 21 CFR 820.22 | Quality audits - Lack of or inadequate procedures |
| 2350 | 21 CFR 820.25(b) | Training - Lack of or inadequate procedures |
| 3101 | 21 CFR 820.30(g) | Design validation- Lack of or inadequate procedures |
| 3117 | 21 CFR 820.70(i) | Software validation for automated processes |
| 3130 | 21 CFR 820.100(a) | Lack of or inadequate procedures |
| 3160 | 21 CFR 820.184 | Lack of or inadequate DHR procedures |
| 3345 | 21 CFR 820.200(a) | Servicing - Lack of or inadequate procedures |
| 3666 | 21 CFR 820.20(c) | Management review - Lack of or inadequate procedures |
| 479 | 21 CFR 820.50 | Purchasing controls, Lack of or inadequate procedures |
| 5008 | 21 CFR 1002.20(a) | Failure to report |
| 5034 | 21 CFR 1010.2(c) | Certification not based on adequate test/testing program |
| 546 | 21 CFR 820.75(a) | Lack of or inadequate process validation |
| 630 | 21 CFR 803.17 | Lack of Written MDR Procedures |