1025529

Huestis Machine Corporation — FEI 1219183

Inspection Details

Classification
Voluntary Action Indicated (VAI) (VAI)
End Date
September 15, 2017
Fiscal Year
2017
Product Type
Devices
Project Area
Postmarket Assurance: Devices
Location
Bristol, RI (United States)

Citations

IDCFRDescription
1471321 CFR 820.198(a)Lack of or inadequate complaint procedures
310321 CFR 820.30(i)Design changes - Lack of or Inadequate Procedures
311721 CFR 820.70(i)Software validation for automated processes
312021 CFR 820.80(a)Lack of or inadequate procedures - Acceptance activities
313021 CFR 820.100(a)Lack of or inadequate procedures
328221 CFR 820.90(a)Nonconforming product, Lack of or inadequate procedures
367721 CFR 820.30(g)Design validation - software validation not performed
47921 CFR 820.50Purchasing controls, Lack of or inadequate procedures
503421 CFR 1010.2(c)Certification not based on adequate test/testing program
54621 CFR 820.75(a)Lack of or inadequate process validation
63021 CFR 803.17Lack of Written MDR Procedures