1074875

Steiner Biotechnology LLC — FEI 3005138409

Inspection Details

Classification
Voluntary Action Indicated (VAI) (VAI)
End Date
December 13, 2018
Fiscal Year
2019
Product Type
Devices
Project Area
Compliance: Devices
Location
Henderson, NV (United States)

Citations

IDCFRDescription
1471321 CFR 820.198(a)Lack of or inadequate complaint procedures
232721 CFR 820.22Quality audits - Lack of or inadequate procedures
237121 CFR 820.30(a)Design control - no procedures
312121 CFR 820.80(b)Lack of or inadequate receiving acceptance procedures
313221 CFR 820.120Lack of or inadequate procedures for labeling
316021 CFR 820.184Lack of or inadequate DHR procedures
317221 CFR 820.198(c)Investigation of device failures
366621 CFR 820.20(c)Management review - Lack of or inadequate procedures
367821 CFR 820.30(g)Design Validation - Risk analysis not performed/inadequate
47921 CFR 820.50Purchasing controls, Lack of or inadequate procedures
54621 CFR 820.75(a)Lack of or inadequate process validation