1078904

Lusys Laboratories, Inc. — FEI 3004473533

Inspection Details

Classification
Voluntary Action Indicated (VAI) (VAI)
End Date
February 7, 2019
Fiscal Year
2019
Product Type
Devices
Project Area
Compliance: Devices
Location
San Diego, CA (United States)

Citations

IDCFRDescription
1471121 CFR 820.160(a)Lack of or inadequate procedures for distribution
1471221 CFR 820.184DHR - not or inadequately maintained
1472021 CFR 820.50(a)(3)Acceptable supplier records, inadequate records
1472221 CFR 820.40Procedures not adequately established or maintained
232721 CFR 820.22Quality audits - Lack of or inadequate procedures
310121 CFR 820.30(g)Design validation- Lack of or inadequate procedures
312021 CFR 820.80(a)Lack of or inadequate procedures - Acceptance activities
313021 CFR 820.100(a)Lack of or inadequate procedures
323321 CFR 820.72(a)Calibration, Inspection, etc. Procedures Lack of or Inadequ
323521 CFR 820.72(a)Equipment control activity documentation
326221 CFR 820.250(a)Statistical techniques - Lack of or inadequate procedures
328221 CFR 820.90(a)Nonconforming product, Lack of or inadequate procedures
333121 CFR 820.181DMR - not or inadequately maintained
366621 CFR 820.20(c)Management review - Lack of or inadequate procedures
367821 CFR 820.30(g)Design Validation - Risk analysis not performed/inadequate
368321 CFR 820.70(g)Equipment Installation, Placement, Specified Requirements
383721 CFR 820.25(b)Training records
54121 CFR 820.70(c)Environmental control Lack of or inadequate procedures