FDA Inspection 1100385 — USHIO GERMANY GmbH

USHIO GERMANY GmbH — FEI 3425

The FDA completed an inspection of USHIO GERMANY GmbH on August 23, 2019 in Steinhoring. The inspection was classified OAI (Official Action Indicated — objectionable conditions were found and regulatory action is recommended). 6 citations were issued during this inspection. Product area: Devices.

Inspection Details

Classification
Official Action Indicated (OAI) (OAI)
End Date
August 23, 2019
Fiscal Year
2019
Product Type
Devices
Project Area
Compliance: Devices
Location
Steinhoring (Germany)

Citations

IDCFRDescription
312321 CFR 820.80(c)Lack of or inadequate In-process acceptance procedures
313021 CFR 820.100(a)Lack of or inadequate procedures
315921 CFR 820.184DHR content
326221 CFR 820.250(a)Statistical techniques - Lack of or inadequate procedures
328221 CFR 820.90(a)Nonconforming product, Lack of or inadequate procedures
54621 CFR 820.75(a)Lack of or inadequate process validation