FDA Inspection 1238217 — Criticare Technologies, Inc.

Criticare Technologies, Inc. — FEI 3012238587

The FDA completed an inspection of Criticare Technologies, Inc. on May 10, 2024 in Warwick, RI. The inspection was classified OAI (Official Action Indicated — objectionable conditions were found and regulatory action is recommended). 10 citations were issued during this inspection. Product area: Devices.

Inspection Details

Classification
Official Action Indicated (OAI) (OAI)
End Date
May 10, 2024
Fiscal Year
2024
Product Type
Devices
Project Area
Compliance: Devices
Location
Warwick, RI (United States)

Citations

IDCFRDescription
1471321 CFR 820.198(a)Lack of or inadequate complaint procedures
1471621 CFR 820.30(f)Design verification - output does not meet input requirement
2280021 CFR 830.300(a)Devices subject to device identification GUDID data submission requirements.
310121 CFR 820.30(g)Design validation- Lack of or inadequate procedures
312021 CFR 820.80(a)Lack of or inadequate procedures - Acceptance activities
313021 CFR 820.100(a)Lack of or inadequate procedures
328221 CFR 820.90(a)Nonconforming product, Lack of or inadequate procedures
367821 CFR 820.30(g)Design Validation - Risk analysis not performed/inadequate
383721 CFR 820.25(b)Training records
47921 CFR 820.50Purchasing controls, Lack of or inadequate procedures