1243932

Phillips Precision, Inc. — FEI 3003387384

Inspection Details

Classification
Official Action Indicated (OAI) (OAI)
End Date
July 17, 2024
Fiscal Year
2024
Product Type
Devices
Project Area
Compliance: Devices
Location
Elmwood Park, NJ (United States)

Citations

IDCFRDescription
235021 CFR 820.25(b)Training - Lack of or inadequate procedures
310821 CFR 820.70(e)Contamination control, Lack of or inadequate procedures
311921 CFR 820.75(b)Lack/Inad procedure-Monitoring/Control of Validated Proces
312021 CFR 820.80(a)Lack of or inadequate procedures - Acceptance activities
313021 CFR 820.100(a)Lack of or inadequate procedures
322621 CFR 820.70(g)(1)Maintenance schedule, Lack of or inadequate schedule
328221 CFR 820.90(a)Nonconforming product, Lack of or inadequate procedures
54621 CFR 820.75(a)Lack of or inadequate process validation