FDA Inspection 1260616 — Audifon GmbH & Co. KG

Audifon GmbH & Co. KG — FEI 3006076696

The FDA completed an inspection of Audifon GmbH & Co. KG on Jan 16, 2025 in Kolleda. The inspection was classified VAI (Voluntary Action Indicated — objectionable conditions or practices were found, but the agency is not prepared to take or recommend any administrative or regulatory action). 6 citations were issued during this inspection. Product area: Devices.

Inspection details

Classification
Voluntary Action Indicated (VAI)
End Date
Jan 16, 2025
Fiscal Year
2025
Product Type
Devices
Project Area
Compliance: Devices
Location
Kolleda (Germany)

Citations

IDCFRDescription
1471321 CFR 820.198(a)Lack of or inadequate complaint procedures
310121 CFR 820.30(g)Design validation- Lack of or inadequate procedures
310321 CFR 820.30(i)Design changes - Lack of or Inadequate Procedures
313021 CFR 820.100(a)Lack of or inadequate procedures
322621 CFR 820.70(g)(1)Maintenance schedule, Lack of or inadequate schedule
47921 CFR 820.50Purchasing controls, Lack of or inadequate procedures

Need the full FDA record for Audifon GmbH & Co. KG? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Audifon GmbH & Co. KG: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Audifon GmbH & Co. KG →

Sources: openFDA, FDA Data Dashboard, FDA.gov. Report a data issue: hello@regdatalab.com.