FDA Inspection 728178 — Archtek Dental, LLC

Archtek Dental, LLC — FEI 3000203342

The FDA completed an inspection of Archtek Dental, LLC on June 2, 2011 in Carmel, IN. The inspection was classified OAI (Official Action Indicated — objectionable conditions were found and regulatory action is recommended). 15 citations were issued during this inspection. Product area: Devices.

Inspection Details

Classification
Official Action Indicated (OAI) (OAI)
End Date
June 2, 2011
Fiscal Year
2011
Product Type
Devices
Project Area
Compliance: Devices
Location
Carmel, IN (United States)

Citations

IDCFRDescription
1471321 CFR 820.198(a)Lack of or inadequate complaint procedures
230221 CFR 820.20(e)Quality System Procedures
232721 CFR 820.22Quality audits - Lack of or inadequate procedures
235021 CFR 820.25(b)Training - Lack of or inadequate procedures
312021 CFR 820.80(a)Lack of or inadequate procedures - Acceptance activities
313021 CFR 820.100(a)Lack of or inadequate procedures
316021 CFR 820.184Lack of or inadequate DHR procedures
316821 CFR 820.198(a)Complaints
322621 CFR 820.70(g)(1)Maintenance schedule, Lack of or inadequate schedule
328221 CFR 820.90(a)Nonconforming product, Lack of or inadequate procedures
333121 CFR 820.181DMR - not or inadequately maintained
368021 CFR 820.70(a)Process control procedures, Lack of or inadequate procedures
47921 CFR 820.50Purchasing controls, Lack of or inadequate procedures
54621 CFR 820.75(a)Lack of or inadequate process validation
63021 CFR 803.17Lack of Written MDR Procedures