728640

DePuy Orthopaedics, Inc. — FEI 1818910

Inspection Details

Classification
Official Action Indicated (OAI) (OAI)
End Date
June 7, 2011
Fiscal Year
2011
Product Type
Devices
Project Area
Compliance: Devices
Location
Warsaw, IN (United States)

Citations

IDCFRDescription
1471321 CFR 820.198(a)Lack of or inadequate complaint procedures
1471821 CFR 820.30(g)Design validation - Risk analysis
312321 CFR 820.80(c)Lack of or inadequate In-process acceptance procedures
313021 CFR 820.100(a)Lack of or inadequate procedures
317221 CFR 820.198(c)Investigation of device failures
328221 CFR 820.90(a)Nonconforming product, Lack of or inadequate procedures