FDA Inspection 767748 — NeuroVasix, Inc.

NeuroVasix, Inc. — FEI 3006510104

The FDA completed an inspection of NeuroVasix, Inc. on February 15, 2012 in Broken Arrow, OK. The inspection was classified OAI (Official Action Indicated — objectionable conditions were found and regulatory action is recommended). 10 citations were issued during this inspection. Product area: Devices.

Inspection Details

Classification
Official Action Indicated (OAI) (OAI)
End Date
February 15, 2012
Fiscal Year
2012
Product Type
Devices
Project Area
Postmarket Assurance: Devices
Location
Broken Arrow, OK (United States)

Citations

IDCFRDescription
1471321 CFR 820.198(a)Lack of or inadequate complaint procedures
237121 CFR 820.30(a)Design control - no procedures
310421 CFR 820.30(j)Design history file
312121 CFR 820.80(b)Lack of or inadequate receiving acceptance procedures
316021 CFR 820.184Lack of or inadequate DHR procedures
317221 CFR 820.198(c)Investigation of device failures
333121 CFR 820.181DMR - not or inadequately maintained
405921 CFR 820.22Quality Audits - defined intervals
47921 CFR 820.50Purchasing controls, Lack of or inadequate procedures
63021 CFR 803.17Lack of Written MDR Procedures