FDA Inspection 780737 — MEDDEV USA, INC.

MEDDEV USA, INC. — FEI 3009257101

The FDA completed an inspection of MEDDEV USA, INC. on May 11, 2012 in Dallas, TX. The inspection was classified OAI (Official Action Indicated — objectionable conditions were found and regulatory action is recommended). 10 citations were issued during this inspection. Product area: Devices.

Inspection Details

Classification
Official Action Indicated (OAI) (OAI)
End Date
May 11, 2012
Fiscal Year
2012
Product Type
Devices
Project Area
Compliance: Devices
Location
Dallas, TX (United States)

Citations

IDCFRDescription
1471321 CFR 820.198(a)Lack of or inadequate complaint procedures
226921 CFR 820.20(a)Quality policy and objectives
237121 CFR 820.30(a)Design control - no procedures
312021 CFR 820.80(a)Lack of or inadequate procedures - Acceptance activities
313021 CFR 820.100(a)Lack of or inadequate procedures
316021 CFR 820.184Lack of or inadequate DHR procedures
333121 CFR 820.181DMR - not or inadequately maintained
367821 CFR 820.30(g)Design Validation - Risk analysis not performed/inadequate
47921 CFR 820.50Purchasing controls, Lack of or inadequate procedures
63021 CFR 803.17Lack of Written MDR Procedures