The FDA completed an inspection of LivaNova USA, Inc. on Dec 7, 2012 in Houston, TX. The inspection was classified VAI (Voluntary Action Indicated — objectionable conditions or practices were found, but the agency is not prepared to take or recommend any administrative or regulatory action). 10 citations were issued during this inspection. Product area: Devices.

Inspection details

Classification
Voluntary Action Indicated (VAI)
End Date
Dec 7, 2012
Fiscal Year
2013
Product Type
Devices
Project Area
Compliance: Devices
Location
Houston, TX (United States)

Citations

IDCFRDescription
1471321 CFR 820.198(a)Lack of or inadequate complaint procedures
1471921 CFR 820.30(h)Incorrect translation to production specifications
312821 CFR 820.90(a)Nonconforming product control
313021 CFR 820.100(a)Lack of or inadequate procedures
316821 CFR 820.198(a)Complaints
328521 CFR 820.90(b)(2)Product rework procedures, Lack of or inadequate procedures
367621 CFR 820.30(f)Design verification - documentation
368021 CFR 820.70(a)Process control procedures, Lack of or inadequate procedures
54621 CFR 820.75(a)Lack of or inadequate process validation
73121 CFR 803.50(a)(1)Report of Death or Serious Injury