FDA Inspection 889188 — Simply Clean Air & Water Inc.

Simply Clean Air & Water Inc. — FEI 3004114902

The FDA completed an inspection of Simply Clean Air & Water Inc. on July 18, 2014 in Newington, CT. The inspection was classified OAI (Official Action Indicated — objectionable conditions were found and regulatory action is recommended). 12 citations were issued during this inspection. Product area: Devices.

Inspection Details

Classification
Official Action Indicated (OAI) (OAI)
End Date
July 18, 2014
Fiscal Year
2014
Product Type
Devices
Project Area
Compliance: Devices
Location
Newington, CT (United States)

Citations

IDCFRDescription
232821 CFR 820.22Quality audits - auditor independence
237121 CFR 820.30(a)Design control - no procedures
310321 CFR 820.30(i)Design changes - Lack of or Inadequate Procedures
312021 CFR 820.80(a)Lack of or inadequate procedures - Acceptance activities
315921 CFR 820.184DHR content
319321 CFR 820.30(g)Design validation - simulated testing
323521 CFR 820.72(a)Equipment control activity documentation
323621 CFR 820.72(b)Calibration procedures - content
332321 CFR 820.170(b)Installer records
333121 CFR 820.181DMR - not or inadequately maintained
367121 CFR 820.25(a)Personnel
47921 CFR 820.50Purchasing controls, Lack of or inadequate procedures