896112
U.S. Infusion, LLC — FEI 3008782867
Inspection Details
- Classification
- Official Action Indicated (OAI) (OAI)
- End Date
- September 17, 2014
- Fiscal Year
- 2014
- Product Type
- Devices
- Project Area
- Compliance: Devices
- Location
- South Miami, FL (United States)
Citations
| ID | CFR | Description |
| 14713 | 21 CFR 820.198(a) | Lack of or inadequate complaint procedures |
| 14722 | 21 CFR 820.40 | Procedures not adequately established or maintained |
| 2327 | 21 CFR 820.22 | Quality audits - Lack of or inadequate procedures |
| 3101 | 21 CFR 820.30(g) | Design validation- Lack of or inadequate procedures |
| 3103 | 21 CFR 820.30(i) | Design changes - Lack of or Inadequate Procedures |
| 3130 | 21 CFR 820.100(a) | Lack of or inadequate procedures |
| 3696 | 21 CFR 820.100(b) | Documentation |
| 479 | 21 CFR 820.50 | Purchasing controls, Lack of or inadequate procedures |
| 630 | 21 CFR 803.17 | Lack of Written MDR Procedures |
| 732 | 21 CFR 803.50(a)(2) | Individual Report of Malfunction |