FDA Inspection 923313 — Shanghai Gaohui Rubber And Plastic Company
The FDA completed an inspection of Shanghai Gaohui Rubber And Plastic Company on March 26, 2015 in Shanghai. The inspection was classified OAI (Official Action Indicated — objectionable conditions were found and regulatory action is recommended). 13 citations were issued during this inspection. Product area: Devices.
Inspection Details
- Classification
- Official Action Indicated (OAI) (OAI)
- End Date
- March 26, 2015
- Fiscal Year
- 2015
- Product Type
- Devices
- Project Area
- Compliance: Devices
- Location
- Shanghai (China)
Citations
| ID | CFR | Description |
|---|---|---|
| 14712 | 21 CFR 820.184 | DHR - not or inadequately maintained |
| 14713 | 21 CFR 820.198(a) | Lack of or inadequate complaint procedures |
| 2327 | 21 CFR 820.22 | Quality audits - Lack of or inadequate procedures |
| 2604 | 21 CFR 820.30(e) | Design review - Lack of or inadequate procedures |
| 2650 | 21 CFR 820.30(f) | Design verification - Lack of or inadequate procedures |
| 3101 | 21 CFR 820.30(g) | Design validation- Lack of or inadequate procedures |
| 3104 | 21 CFR 820.30(j) | Design history file |
| 3118 | 21 CFR 820.75(a) | Documentation |
| 3130 | 21 CFR 820.100(a) | Lack of or inadequate procedures |
| 3235 | 21 CFR 820.72(a) | Equipment control activity documentation |
| 3262 | 21 CFR 820.250(a) | Statistical techniques - Lack of or inadequate procedures |
| 3286 | 21 CFR 820.90(b)(1) | Procedures for product review,disposition lack of/inadequate |
| 631 | 21 CFR 803.17(a)(1) | Lack of System for Event Evaluations |