FDA Inspection 949799 — Helica Instruments, Ltd.

Helica Instruments, Ltd. — FEI 3003824921

The FDA completed an inspection of Helica Instruments, Ltd. on November 11, 2015 in Edinburgh. The inspection was classified OAI (Official Action Indicated — objectionable conditions were found and regulatory action is recommended). 12 citations were issued during this inspection. Product area: Devices.

Inspection Details

Classification
Official Action Indicated (OAI) (OAI)
End Date
November 11, 2015
Fiscal Year
2016
Product Type
Devices
Project Area
Postmarket Assurance: Devices
Location
Edinburgh (United Kingdom)

Citations

IDCFRDescription
1471321 CFR 820.198(a)Lack of or inadequate complaint procedures
1471421 CFR 820.30(c)Design input - Lack of or inadequate procedures
260421 CFR 820.30(e)Design review - Lack of or inadequate procedures
265021 CFR 820.30(f)Design verification - Lack of or inadequate procedures
310121 CFR 820.30(g)Design validation- Lack of or inadequate procedures
310221 CFR 820.30(h)Design transfer - Lack of or inadequate procedures
310321 CFR 820.30(i)Design changes - Lack of or Inadequate Procedures
312521 CFR 820.80(d)Lack of or inadequate final acceptance procedures
313021 CFR 820.100(a)Lack of or inadequate procedures
328221 CFR 820.90(a)Nonconforming product, Lack of or inadequate procedures
47921 CFR 820.50Purchasing controls, Lack of or inadequate procedures
63021 CFR 803.17Lack of Written MDR Procedures