FDA Inspection 959637 — Spiegelberg Gmbh & Co. KG

Spiegelberg Gmbh & Co. KG — FEI 3002108226

The FDA completed an inspection of Spiegelberg Gmbh & Co. KG on February 4, 2016 in Hamburg. The inspection was classified OAI (Official Action Indicated — objectionable conditions were found and regulatory action is recommended). 12 citations were issued during this inspection. Product area: Devices.

Inspection Details

Classification
Official Action Indicated (OAI) (OAI)
End Date
February 4, 2016
Fiscal Year
2016
Product Type
Devices
Project Area
Compliance: Devices
Location
Hamburg (Germany)

Citations

IDCFRDescription
1471321 CFR 820.198(a)Lack of or inadequate complaint procedures
310121 CFR 820.30(g)Design validation- Lack of or inadequate procedures
310321 CFR 820.30(i)Design changes - Lack of or Inadequate Procedures
311721 CFR 820.70(i)Software validation for automated processes
313021 CFR 820.100(a)Lack of or inadequate procedures
326421 CFR 820.250(b)Sampling methods - Lack of or inadequate procedures
328221 CFR 820.90(a)Nonconforming product, Lack of or inadequate procedures
419121 CFR 806.10(a)(1)Report of risk to health
47921 CFR 820.50Purchasing controls, Lack of or inadequate procedures
63021 CFR 803.17Lack of Written MDR Procedures
73221 CFR 803.50(a)(2)Individual Report of Malfunction
79321 CFR 803.53(a)Failure to Submit 5 Day Report