981584
Zyno Medical LLC — FEI 3006575795
Inspection Details
- Classification
- Official Action Indicated (OAI) (OAI)
- End Date
- August 11, 2016
- Fiscal Year
- 2016
- Product Type
- Devices
- Project Area
- Compliance: Devices
- Location
- Natick, MA (United States)
Citations
| ID | CFR | Description |
| 14713 | 21 CFR 820.198(a) | Lack of or inadequate complaint procedures |
| 3120 | 21 CFR 820.80(a) | Lack of or inadequate procedures - Acceptance activities |
| 3130 | 21 CFR 820.100(a) | Lack of or inadequate procedures |
| 3159 | 21 CFR 820.184 | DHR content |
| 3192 | 21 CFR 820.30(g) | Design validation - user needs and intended uses |
| 3282 | 21 CFR 820.90(a) | Nonconforming product, Lack of or inadequate procedures |
| 3345 | 21 CFR 820.200(a) | Servicing - Lack of or inadequate procedures |
| 732 | 21 CFR 803.50(a)(2) | Individual Report of Malfunction |