Z-0001-2017 Class III Terminated

Recalled by Terumo Cardiovascular Systems Corporation — Elkton, MD

FDA device recall Z-0001-2017 was initiated by Terumo Cardiovascular Systems Corporation on Dec 11, 2015 and is designated Class III. Reason for recall: Terumo Cardiovascular Systems (Terumo CVS) has received complaints of the CDI System 500 monitor displaying the H/S DISCONNECT AT CUVETTE error message when the CDI H/S Cuvette does not make a… The recall status is terminated (terminated Oct 26, 2016). Affected quantity: 140,000 units.

Recall Details

Product Type
Devices
Report Date
Oct 12, 2016
Initiation Date
Dec 11, 2015
Termination Date
Oct 26, 2016
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
140,000 units

Product Description

CDI¿ H/S Cuvette The intended use for the CDI¿ H IS Cuvette is as follows: The CDI¿ H/S Cuvettes are intended for use with a CDI monitor during cardiopulmonary bypass procedures when continuous monitoring of blood hematocrit and oxygen saturation is desired.

Reason for Recall

Terumo Cardiovascular Systems (Terumo CVS) has received complaints of the CDI System 500 monitor displaying the H/S DISCONNECT AT CUVETTE error message when the CDI H/S Cuvette does not make a proper connection to the CDI H/S Probe. In this situation, blood parameter values for HCT, Hgb, and SO2 do not display.

Distribution Pattern

Worldwide Distribution - USA (nationwide) Distribution.

Code Information

6912 , 6913 6914, 6932, 6933 6934, 6922, 6923 6924, CV-6912, CV-6913, CV-6914 Cardiovascular Procedure Kits- Various

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0001-2017
openFDA file date
Oct 23, 2026