Z-0002-2018 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- October 11, 2017
- Initiation Date
- August 23, 2017
- Termination Date
- February 21, 2019
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 361 kits
Product Description
Human Alpha-1-Antitrypsin Kit for use on the SPAPlus analyser, Product #NK034.S. The firm name on the label is The Binding Site Group Ltd., Birmingham, UK. Product is designed for the quantitative in-vitro determination of Alpha-1-antitrypsin in human serum using the SPAPLUS turbidimetric analyser. The measurement of Alpha-1-antitrypsin aids in the diagnosis of several conditions including adult cirrhosis of the liver. In addition, Alpha-1-antitrypsin deficiency has been associated with pulmonary emphysema. This test should be used in conjunction with other laboratory and clinical findings.
Reason for Recall
The high control for the kit is recovering low and may be out of range for assigned values.
Distribution Pattern
Worldwide Distribution - US Distribution to the state of CA., and to the countries of : United Kingdom, Germany, France, Czech Republic, Belgium, Panama, Bulgaria, Iran, Argentina, Turkey Romania, Colombia, and Saudi Arabia.
Code Information
Lot #400900, Exp. 12/31/2017, UDI #05051700016421