Z-0003-2017 Class II Terminated

Recalled by Zimmer Biomet, Inc. — Warsaw, IN

FDA device recall Z-0003-2017 was initiated by Zimmer Biomet, Inc. on Aug 8, 2016 and is designated Class II. Reason for recall: The drill bit used to drill the hex pre-drilled hole in the screw head of the Tibial Augment Bolts included in the Vanguard 360 Revision Knee System was larger than allowable per the process… The recall status is terminated (terminated Jun 5, 2017). Affected quantity: 454.

Recall Details

Product Type
Devices
Report Date
Oct 12, 2016
Initiation Date
Aug 8, 2016
Termination Date
Jun 5, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
454

Product Description

Vanguard 360 Revision Knee System Knee prostheses

Reason for Recall

The drill bit used to drill the hex pre-drilled hole in the screw head of the Tibial Augment Bolts included in the Vanguard 360 Revision Knee System was larger than allowable per the process specification. As a result, the hex geometry can still be visible, but there are radial cuts into the hex flat. Therefore, when the bolt is inserted into the augment, the screwdriver is unable to catch on the hex geometry, or the screwdriver strips the remaining hex geometry, and the bolt is unable to be fully tightened. The bolt stripping or the bolt not being fully tightened will be noticed by the surgeon intraoperatively when he or she is unable to screw the augment bolts into the augment. This possible event can result in an intraoperative delay resulting in increased time under anesthesia.

Distribution Pattern

Worldwide Distribution - US Distribution to the states of : VA, UT, TX. OR, MN, AL, PA, CA and KY., and to the countries of : AUSTRALIA, China, Italy, Netherlands and Japan.

Code Information

ITEM LOT DESCRIPTION 185220 262690 BMT 360 TIB AUG 59X5MM 185221 475560 BMT 360 TIB AUG 63X5MM 185222 262720 BMT 360 TIB AUG 67X5MM 185222 475570 BMT 360 TIB AUG 67X5MM 185222 524160 BMT 360 TIB AUG 67X5MM 185222 690770 BMT 360 TIB AUG 67X5MM 185223 421690 BMT 360 TIB AUG 71X5MM 185223 475580 BMT 360 TIB AUG 71X5MM 185223 475590 BMT 360 TIB AUG 71X5MM 185223 524170 BMT 360 TIB AUG 71X5MM 185224 475600 BMT 360 TIB AUG 75X5MM 185224 475610 BMT 360 TIB AUG 75X5MM 185224 524180 BMT 360 TIB AUG 75X5MM 185224 647840 BMT 360 TIB AUG 75X5MM 185224 690790 BMT 360 TIB AUG 75X5MM 185225 342740 BMT 360 TIB AUG 79X5MM 185225 421700 BMT 360 TIB AUG 79X5MM 185225 524190 BMT 360 TIB AUG 79X5MM 185225 524200 BMT 360 TIB AUG 79X5MM 185225 690800 BMT 360 TIB AUG 79X5MM 185226 647850 BMT 360 TIB AUG 83X5MM 185226 751570 BMT 360 TIB AUG 83X5MM 185228 647910 BMT 360 TIB AUG 91X5MM 185228 751420 BMT 360 TIB AUG 91X5MM 185230 524210 BMT 360 TIB AUG 59X10MM 185231 421730 BMT 360 TIB AUG 63X10MM 185231 475620 BMT 360 TIB AUG 63X10MM 185231 524220 BMT 360 TIB AUG 63X10MM 185232 377450 BMT 360 TIB AUG 67X10MM 185232 377460 BMT 360 TIB AUG 67X10MM 185232 421750 BMT 360 TIB AUG 67X10MM 185232 475630 BMT 360 TIB AUG 67X10MM 185232 524230 BMT 360 TIB AUG 67X10MM 185233 342750 BMT 360 TIB AUG 71X10MM 185233 377470 BMT 360 TIB AUG 71X10MM 185233 475640 BMT 360 TIB AUG 71X10MM 185233 524240 BMT 360 TIB AUG 71X10MM 185234 262780 BMT 360 TIB AUG 75X10MM 185234 262790 BMT 360 TIB AUG 75X10MM 185234 524250 BMT 360 TIB AUG 75X10MM 185235 421770 BMT 360 TIB AUG 79X10MM 185235 421780 BMT 360 TIB AUG 79X10MM 185237 229750 BMT 360 TIB AUG 87X10MM 185238 229760 BMT 360 TIB AUG 91X10MM 185238 262820 BMT 360 TIB AUG 91X10MM 185239 511390 BMT 360 TIB AUG BOLT 10MM 185241 148190 BMT 360 TIB AUG 63X15MM RL/LM 185242 969710 BMT 360 TIB AUG 67X15MM RL/LM 185243 377480 BMT 360 TIB AUG 71X15MM RL/LM 185247 262750 BMT 360 TIB AUG 87X15 RL/LM 185250 421790 BMT 360 TIB AUG 59X15MM LL/RM 185251 262770 BMT 360 TIB AUG 63X15MM LL/RM 185253 421800 BMT 360 TIB AUG 71X15MM LL/RM 185254 229770 BMT 360 TIB AUG 75X15MM LL/RM 185255 229780 BMT 360 TIB AUG 79X15MM LL/RM 185256 988680 BMT 360 TIB AUG 83X15MM LL/RM 185257 148210 BMT 360 TIB AUG 87X15 LL/RM 185229-00 493640 BMT TIB AUG BOLT 5MM 185229-00 582090 BMT TIB AUG BOLT 5MM 185229-00 582100 BMT TIB AUG BOLT 5MM 185229-00 582110 BMT TIB AUG BOLT 5MM 185229-00 590570 BMT TIB AUG BOLT 5MM 185229-00 789260 BMT TIB AUG BOLT 5MM 185239-00 493690 BMT 360 TIB AUG BOLT 10MM 185239-00 590580 BMT 360 TIB AUG BOLT 10MM 185239-00 590590 BMT 360 TIB AUG BOLT 10MM 185239-00 590600 BMT 360 TIB AUG BOLT 10MM 185239-00 590610 BMT 360 TIB AUG BOLT 10MM 185239-00 590620 BMT 360 TIB AUG BOLT 10MM 185239-00 590630 BMT 360 TIB AUG BOLT 10MM 185239-00 666760 BMT 360 TIB AUG BOLT 10MM 185239-00 666780 BMT 360 TIB AUG BOLT 10MM

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0003-2017
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.