Z-0013-2025 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- October 30, 2024
- Initiation Date
- June 26, 2024
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 119535 units
Product Description
CADD Medication Cassette Reservoir, 50mL, FS, Product Code 21-7301-24
Reason for Recall
Certain CADD Medication Cassette Reservoirs may exhibit a weakened weld joint between the medication bag and tubing due to a production equipment malfunction. This could result in a medication leakage.
Distribution Pattern
Worldwide distribution.
Code Information
UDI/DI 10610586027215, Lot Numbers: 4042839, 4040248, 4034021, 4052405, 4034022, 4037748, 4042841, 4034019, 4037749, 4040247, 3983339, 3983338, 3983340, 4013367, 4013366, 3988532-ENG, 3983337, 4052406, 4042840, 4052407, 4052408, 4042842, 4048891, 4052409