Z-0013-2025 Class II Ongoing

Recalled by Smiths Medical ASD Inc. — Minneapolis, MN

FDA device recall Z-0013-2025 was initiated by Smiths Medical ASD Inc. on Jun 26, 2024 and is designated Class II. Reason for recall: Certain CADD Medication Cassette Reservoirs may exhibit a weakened weld joint between the medication bag and tubing due to a production equipment malfunction. This could result in a medication… The recall status is ongoing. Affected quantity: 119535 units.

Recall Details

Product Type
Devices
Report Date
Oct 30, 2024
Initiation Date
Jun 26, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
119535 units

Product Description

CADD Medication Cassette Reservoir, 50mL, FS, Product Code 21-7301-24

Reason for Recall

Certain CADD Medication Cassette Reservoirs may exhibit a weakened weld joint between the medication bag and tubing due to a production equipment malfunction. This could result in a medication leakage.

Distribution Pattern

Worldwide distribution.

Code Information

UDI/DI 10610586027215, Lot Numbers: 4042839, 4040248, 4034021, 4052405, 4034022, 4037748, 4042841, 4034019, 4037749, 4040247, 3983339, 3983338, 3983340, 4013367, 4013366, 3988532-ENG, 3983337, 4052406, 4042840, 4052407, 4052408, 4042842, 4048891, 4052409

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0013-2025
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.