Z-0015-2025 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- October 30, 2024
- Initiation Date
- June 26, 2024
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 109563 units
Product Description
CADD Medication Cassette Reservoir, 100mL, FS, Product Code 21-7302-24JP
Reason for Recall
Certain CADD Medication Cassette Reservoirs may exhibit a weakened weld joint between the medication bag and tubing due to a production equipment malfunction. This could result in a medication leakage.
Distribution Pattern
Worldwide distribution.
Code Information
UID/DI NA, Lot/Serial Numbers: 3983315, 3971709, 3983316, 3983320, 3983329, 4013361, 4013363, 4013373, 4013377, 4013378, 4019427, 4019429, 4034028, 4037753, 4052410, 4040250, 4042843, 4048894, 4046846, 4052412, 4052413