Z-0030-2020 Class II Terminated

Recalled by Baxter Healthcare Corporation — Deerfield, IL

Recall Details

Product Type
Devices
Report Date
October 9, 2019
Initiation Date
October 10, 2018
Termination Date
July 22, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
8640 eaches

Product Description

L5C4531 X1 Integrated APD Set with Cassette 3-Prong Sterile, nonpyrogenic fluid path. Sterilized by ethylene oxide.

Reason for Recall

Potential presence of leaks from the cassettes, which may be present due to an incomplete seal of the sheeting to the cassette. A leaking cassette could lead to contamination of the sterile fluid path with micro-organisms. This may predispose the patient to peritonitis.

Distribution Pattern

U.S.: VA, DC, CA, FL, SC, MD, DE, PA, NC, WA, GA, TX, NV. No foreign (OUS) distribution.

Code Information

Lot Number H18D22030