Z-1368-2026 Class II Ongoing
FDA device recall Z-1368-2026 was initiated by Baxter Healthcare Corporation on Jan 13, 2026 and is designated Class II. Reason for recall: Pumps were released without full testing being performed, including occlusion alarm testing. The recall status is ongoing. Affected quantity: 5.
Recall details
- Product Type
- Devices
- Report Date
- Feb 25, 2026
- Initiation Date
- Jan 13, 2026
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 5
Product description
Brand Name: Spectrum IQ Product Name: Infusion Pump Model/Catalog Number: 3570009 Software Version: N/A Product Description: Infusion Pump, one unit per carton Component: N/A
Reason for recall
Pumps were released without full testing being performed, including occlusion alarm testing.
Distribution pattern
States: Ohio Countries: United States
Code information
Lot Code: GTIN 00085412610900, Serial Numbers 3815663, 3823468, 3824743, 3825234, and 3825599