Z-0333-2027 Class II Ongoing

Recalled by Baxter Healthcare Corporation — Round Lake, IL

FDA device recall Z-0333-2027 was initiated by Baxter Healthcare Corporation on Sep 14, 2026 and is designated Class II. Reason for recall: The main lift may unexpectedly drop from its set position to its lowest position during use. The tabletop may not be able to be raised again after the drop occurs. An uncontrolled drop of the surgical… The recall status is ongoing. Affected quantity: 1 unit.

Recall details

Product Type
Devices
Report Date
Nov 11, 2026
Initiation Date
Sep 14, 2026
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1 unit

Product description

Mobile column TruSystem 7500 U, Product Code 1730720

Reason for recall

The main lift may unexpectedly drop from its set position to its lowest position during use. The tabletop may not be able to be raised again after the drop occurs. An uncontrolled drop of the surgical table may result in an abrupt patient displacement during surgery. Serious or critical adverse health consequences may occur, including traumatic injury, airway compromise, and surgical procedure complications, among others.

Distribution pattern

US Nationwide distribution in the states of IL, MO, TX.

Code information

UDI-DI: 00887761974388; Serial number: 110183858

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0333-2027
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Baxter Healthcare Corporation? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-0333-2027. Report a data issue: hello@regdatalab.com.