Z-0352-2027 Class II Ongoing

Recalled by Baxter Healthcare Corporation — Deerfield, IL

FDA device recall Z-0352-2027 was initiated by Baxter Healthcare Corporation on Sep 29, 2026 and is designated Class II. Reason for recall: There is an issue with the main lift mechanism. The main lift may unexpectedly drop from its set position to its lowest position during use. In these cases, the tabletop could not be raised again… The recall status is ongoing. Affected quantity: 16 units.

Recall details

Product Type
Devices
Report Date
Nov 11, 2026
Initiation Date
Sep 29, 2026
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
16 units

Product description

Operating Table TruSystem 7000 U, Surgical table, universal, Model/Catalog Number: 1604788

Reason for recall

There is an issue with the main lift mechanism. The main lift may unexpectedly drop from its set position to its lowest position during use. In these cases, the tabletop could not be raised again after the drop occurred. The issue has been associated with unexpected wear of the main lift assembly, also known as the telescopic spindle. An uncontrolled drop of the surgical table may result in an abrupt patient displacement during surgery. Serious or critical adverse health consequences may occur, including traumatic injury, airway compromise, and surgical procedure complications, among others. Baxter has received three complaints related to this issue, with no reported patient injury.

Distribution pattern

Worldwide - US Nationwide distribution.

Code information

Serial numbers: 108517025, 110115554, 110144247, 110188916, 109558596, 110139748, 110166201, 110188918, 109737582, 110139769, 110177678, 110192686, 110115551, 110144245, 110185616, 109889060; UDI-DI number: 00887761967526

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0352-2027
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.