Z-0041-2026 Class II Ongoing

Recalled by Fresenius Medical Care Holdings, Inc. — Waltham, MA

Recall Details

Product Type
Devices
Report Date
October 15, 2025
Initiation Date
August 29, 2025
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
172,494,732 units (14,374,561 cases @ 12 units per case)

Product Description

Optiflux¿ High Flux E-beam Dialyzer Model Numbers: (1) 0500316E OPTIFLUX F160NRe 12/CS 1.5SA EBEAM, (2) 0500318E OPTIFLUX F180NRe 12/CS 1.7SA EBEAM, (3) 0500320E OPTIFLUX F200NRe 12/CS 1.9SA EBEAM, (4) 0500325E OPTIFLUX F250NRe 12/CS 2.5SA EBEAM

Reason for Recall

To provide further information to the user to ensure proper use of the medical device. the caps had been changed from a threaded version to a press-on version whereby the user must press the caps firmly onto the dialyzer to securely affix the cap.

Distribution Pattern

US Nationwide distribution.

Code Information

1. Model: 0500316E OPTIFLUX F160NRe 12/CS 1.5SA EBEAM, UDI-DI: 840861100149, All Lots manufactured after SEP/06/2023, 2. Model: 0500318E OPTIFLUX F180NRe 12/CS 1.7SA EBEAM, UDI-DI: 840861100156, All Lots manufactured after SEP/06/2023, 3. Model: 0500320E OPTIFLUX F200NRe 12/CS 1.9SA EBEAM, UDI-DI: 840861100163, All Lots manufactured after SEP/06/2023, 4. Model: 0500325E OPTIFLUX F250NRe 12/CS 2.5SA EBEAM, UDI-DI: 840861100170, All Lots manufactured after SEP/06/2023;