Z-2344-2024 Class II Ongoing

Recalled by Fresenius Medical Care Holdings, Inc. — Waltham, MA

FDA device recall Z-2344-2024 was initiated by Fresenius Medical Care Holdings, Inc. on Jun 3, 2024 and is designated Class II. Reason for recall: An increased number of complaints were reported on the Blood Pump Rotor (P/N - F40015481 REV A) after a period of use by clinics noting loose/dislodged guide sheaves. The recall status is ongoing. Affected quantity: 17 Units.

Recall details

Product Type
Devices
Report Date
Jul 17, 2024
Initiation Date
Jun 3, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
17 Units

Product description

RTL190395: 2008K@HOME MACHINE, SHORT CAB,OLC/DP,HP - Blood Pump Rotor spare part, Model Number: F40015481 Rev A

Reason for recall

An increased number of complaints were reported on the Blood Pump Rotor (P/N - F40015481 REV A) after a period of use by clinics noting loose/dislodged guide sheaves.

Distribution pattern

Domestic: Nationwide Distribution. International: Canada.

Code information

Model Number: RTL190395 (Rental Units); UDI/DI: 00840861100958; Serial Numbers: 4K0S-R138, 5K0S-Z652, 5K0S-462B, 5K0S-Z669, 6K0S-505F, 6K0S-513F, 6K0S-850G, 6K0S-480J, 6K0S-467K, 6K0S-995P, 7K0S103784, 6K0S-009M, 7K0S110463, 7K0S110427, 7K0S110771, 6K0S-725R, 6K0S-324L,

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2344-2024
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-2344-2024. Report a data issue: hello@regdatalab.com.