Z-0058-2018 Class II Terminated

Recalled by Biosense Webster, Inc. — Irwindale, CA

Recall Details

Product Type
Devices
Report Date
November 8, 2017
Initiation Date
September 6, 2017
Termination Date
November 19, 2019
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
329 units

Product Description

THERMOCOOL SF NAV Catheters Model Numbers - BNI35FJCT, BNI35DFCT, D-131503-S, D-131504-S

Reason for Recall

Biosense Webster, Inc. has recently received an increased number of complaints related to the display of Alert 402 on the CARTO 3 System for certain lots of THERMOCOOL brand catheters. Alert 402 implies a "Map: magnetic distortion" when connected to CARTO 3 System. This issue may subsequently lead the physician to ablate in an unintended area when delivering RF energy.

Distribution Pattern

US, Japan, Canada, Australia, New Zealand, Brazil, Croatia, Korea, Austria, Colombia, Italy, Portugal, Argentina, South Africa, Belgium, France, Germany, Ireland, Poland, Netherlands, Sweden, Denmark, Finland, Switzerland, Great Britain, Hungary, Czech Republic, China

Code Information

17637495L 17638389L 17639706L 17639707L 17639708L 17640039L 17640494L 17640952L 17640040L 17640742L 17640953L 17640041L 17640743L 17640954L 17640490L 17640744L 17643130L 17638118L 17638387L 17638388L