Z-0064-2019 Class II Terminated

Recalled by Leica Microsystems, Inc. — Buffalo Grove, IL

FDA device recall Z-0064-2019 was initiated by Leica Microsystems, Inc. on Jun 13, 2018 and is designated Class II. Reason for recall: These devices have an incorrect specification against safety standard ISO61010-1. Should a single fault error occur, a mains wire can contact a metal plate which is accessible to the user. This could… The recall status is terminated (terminated Mar 18, 2019). Affected quantity: 67 units.

Recall details

Product Type
Devices
Report Date
Oct 10, 2018
Initiation Date
Jun 13, 2018
Termination Date
Mar 18, 2019
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
67 units

Product description

Leica Biosystems HistoCore SPECTRA ST, Slide Stainer, Automated Product Usage: The intended use of this device is a stainer for laboratory applications and intended for the preparation of histological and cytological specimens.

Reason for recall

These devices have an incorrect specification against safety standard ISO61010-1. Should a single fault error occur, a mains wire can contact a metal plate which is accessible to the user. This could lead to mains power on the metal plate resulting in hazardous voltage to the user.

Distribution pattern

Distributed nationwide to AZ, CA, CO, KS, MA, MH, MO, NJ, OH, WI. Distributed worldwide to CANADA, CHINA, GERMANY, GREAT BRITAIN, JAPAN, SWITZERLAND

Code information

Serial Numbers: 70-78, 111-112, 120-175

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0064-2019
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Leica Microsystems, Inc.? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Leica Microsystems, Inc.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Leica Microsystems, Inc. →

Source: openFDA record for Z-0064-2019. Report a data issue: hello@regdatalab.com.