Z-0064-2019 Class II Terminated

Recalled by Leica Microsystems, Inc. — Buffalo Grove, IL

Recall Details

Product Type
Devices
Report Date
October 10, 2018
Initiation Date
June 13, 2018
Termination Date
March 18, 2019
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
67 units

Product Description

Leica Biosystems HistoCore SPECTRA ST, Slide Stainer, Automated Product Usage: The intended use of this device is a stainer for laboratory applications and intended for the preparation of histological and cytological specimens.

Reason for Recall

These devices have an incorrect specification against safety standard ISO61010-1. Should a single fault error occur, a mains wire can contact a metal plate which is accessible to the user. This could lead to mains power on the metal plate resulting in hazardous voltage to the user.

Distribution Pattern

Distributed nationwide to AZ, CA, CO, KS, MA, MH, MO, NJ, OH, WI. Distributed worldwide to CANADA, CHINA, GERMANY, GREAT BRITAIN, JAPAN, SWITZERLAND

Code Information

Serial Numbers: 70-78, 111-112, 120-175