Z-0069-2019 Class II
FDA device recall Z-0069-2019 was initiated by ICU Medical, Inc. on Aug 20, 2018 and is designated Class II. Reason for recall: Design change implemented changing the device from a blood set with a filter to a transfer set without a filter. Both devices had the same listing number. Customers may not be aware of the product… Affected quantity: 2361 Units.
Recall Details
- Product Type
- Devices
- Report Date
- Oct 10, 2018
- Initiation Date
- Aug 20, 2018
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 2361 Units
Product Description
ICU Medical 14" (36 cm) Appx 3.6 ml, Bifuse Transfer Set w/Rotating Luer, REF: 14237-28 The Intravascular Administration Set is a single use, sterile, non-pyrogenic device which provides access for the administration of fluids from a container to a patient s vascular system through the administration set s needle or catheter (which is inserted into a vein).
Reason for Recall
Design change implemented changing the device from a blood set with a filter to a transfer set without a filter. Both devices had the same listing number. Customers may not be aware of the product design change and use the transfer set as a blood transfusion set without realizing a blood filter is not included.
Distribution Pattern
U.S. Distribution in states of: NY, AL, CA, MS, PA, GA, MI, NC, WI, IL, WA, UT and FL.
Code Information
lots: 85-540-SJ, 86-189-SL, 86-911-SJ