Z-0075-2021 Class II Terminated

Recalled by Biomet 3i, LLC — Palm Beach Gardens, FL

Recall Details

Product Type
Devices
Report Date
October 14, 2020
Initiation Date
August 31, 2020
Termination Date
July 6, 2022
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
9 units

Product Description

CERTAIN BELLATEK TIN ABUTMENT 4.1MM Reference Number: IEDAN4

Reason for Recall

Hex of the BellaTek Encode Definitive Abutments was misaligned by 30 degrees during manufacturing as compared with the provided patient scan data (Images may cause misalignment, the definitive abutment will not seat in the correct orientation upon attempted placement with the patient

Distribution Pattern

FL, GA, ID, IL, MI, NV, NY, OH, SC, TN, VA

Code Information

Lot Numbers: 8563336-1 8563892-1 8563910-1 8565718-2 8571887-1 8571887-2 8571887-3 8572668-1 8572668-2 UDI: (01)00844868031147(10)8563336-1 (01)00844868031147(10)8563892-1 (01)00844868031147(10)8563910-1 (01)00844868031147(10)8565718-2 (01)00844868031147(10)8571887-1 (01)00844868031147(10)8571887-2 (01)00844868031147(10)8571887-3 (01)00844868031147(10)8572668-1 (01)00844868031147(10)8572668-2