Z-0073-2021 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- October 14, 2020
- Initiation Date
- August 31, 2020
- Termination Date
- July 6, 2022
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 7 stand alone units
Product Description
BellaTek Encode Defin3DPMSERitive Abutments-3D Printed Model Reference Numbers: 3DPMSER, 3DPMSERQ
Reason for Recall
Hex of the BellaTek Encode Definitive Abutments was misaligned by 30 degrees during manufacturing as compared with the provided patient scan data (Images may cause misalignment, the definitive abutment will not seat in the correct orientation upon attempted placement with the patient
Distribution Pattern
FL, GA, ID, IL, MI, NV, NY, OH, SC, TN, VA
Code Information
Reference Number: 3DPMSER Lot Numbers: 8560249-1 8564285-1 8570989-1 Reference Number: 3DPMSERQ Lot Numbers: 8561588-1 8561591-1 8569150-1 8570607-1