Z-0076-2013 Class II Terminated

Recalled by The Anspach Effort, Inc. — Palm Beach Gardens, FL

FDA device recall Z-0076-2013 was initiated by The Anspach Effort, Inc. on Jan 9, 2012 and is designated Class II. Reason for recall: The Anspach Effort, Inc. in Palm Beach Gardens, FL is recalling the foot control/foot pedal due to the Foot Control Cable lacking the shield connection from the cable to the connector body. The recall status is terminated (terminated Sep 16, 2014). Affected quantity: 10 units.

Recall Details

Product Type
Devices
Report Date
Oct 24, 2012
Initiation Date
Jan 9, 2012
Termination Date
Sep 16, 2014
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
10 units

Product Description

"***Electric Systems Foot Control with Direction and Irrigation***Rx Only***" Product Usage: Cutting and shaping bone including bones of the spine and cranium.

Reason for Recall

The Anspach Effort, Inc. in Palm Beach Gardens, FL is recalling the foot control/foot pedal due to the Foot Control Cable lacking the shield connection from the cable to the connector body.

Distribution Pattern

Worldwide Distribution - US (nationwide) and the countries of: Italy, Japan, Switzerland, Turkey and Taiwan.

Code Information

Serial Numbers: E3230327501, E32305327502, E32305327503, E32305327504, E32305327505, E32305327601, E32305327602, E32305327603, E32305327604, and E32305327605

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0076-2013
openFDA file date
Oct 26, 2026
Loaded by RegDataLab
Oct 27, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.