Z-0084-2017 Class II Terminated
FDA device recall Z-0084-2017 was initiated by Zimmer Biomet, Inc. on Aug 9, 2016 and is designated Class II. Reason for recall: Old revisions and new revisions of the soft tissue sleeve and recon targeting arm are not interchangeable, as the new revision tissue sleeve will not fit into the old revision targeting arm. The old… The recall status is terminated (terminated Mar 23, 2017). Affected quantity: 31.
Recall Details
- Product Type
- Devices
- Report Date
- Oct 19, 2016
- Initiation Date
- Aug 9, 2016
- Termination Date
- Mar 23, 2017
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 31
Product Description
Phoenix Recon Targeting Arm Guide, Surgical, Instrument for the sleeve and Orthopedic Manual Surgical Instruments for the Arm Product Usage: The Phoenix Femoral Nail System is to be implanted into the femur for alignment, stabilization and fixation of fractures caused by trauma or disease, and the fixation of femurs that have been surgically prepared (osteotomy) for correction of deformity, and for arthrodesis. These instruments are used in support of the surgery.
Reason for Recall
Old revisions and new revisions of the soft tissue sleeve and recon targeting arm are not interchangeable, as the new revision tissue sleeve will not fit into the old revision targeting arm. The old revision tissue sleeve has too much clearance with new revision targeting arm. Risks of encountering the affected product include: the correct instrument may not be readily available for the procedure causing the surgeon to complete the procedure by hand; or a delay in surgery greater than 30 minutes may occur while another Recon Targeting Arm or Recon Soft tissue Sleeve is located or while the surgeon completes the procedure by hand.
Distribution Pattern
US Nationwide Distribution in the states of: MI, UT, SD, OH, WA, KS, MO, NJ, GA, and TX.
Code Information
Item number: 14-442018 Lot number: 290610 290620 386780
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-0084-2017
- openFDA file date
- Oct 23, 2026
- Loaded by RegDataLab