Z-0086-2027 Class II Ongoing

Recalled by Medline Industries, LP — Northfield, IL

FDA device recall Z-0086-2027 was initiated by Medline Industries, LP on Aug 14, 2026 and is designated Class II. Reason for recall: The front fork stem bolt/wheel assembly may disconnect from the rollator frame tubing. If the front fork stem bolt/wheel assembly disconnects from the rollator frame tubing, the front wheels may… The recall status is ongoing. Affected quantity: 1824 units.

Recall Details

Product Type
Devices
Report Date
Oct 14, 2026
Initiation Date
Aug 14, 2026
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1824 units

Product Description

Rollator, Bariatric, Basic Steel, Burgundy, mechanical walker with four legs and a metal frame intended for medical purposes to provide moderate weight support while walking, Product Number MDS86800XWS

Reason for Recall

The front fork stem bolt/wheel assembly may disconnect from the rollator frame tubing. If the front fork stem bolt/wheel assembly disconnects from the rollator frame tubing, the front wheels may become loose or detach from the rollator. If the front wheels become loose or detach during use, there is risk of severe adverse events. The immediate consequences could include falls resulting in head trauma, bone fractures, or other critical injuries.

Distribution Pattern

Worldwide distribution - US Nationwide and the country of Canada.

Code Information

Lot Numbers: 05425090001 05425120001 05426010001 05426020001 05426030001 05426040001 05426050001

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0086-2027
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 24, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.