Z-0089-2023 Class II Ongoing
FDA device recall Z-0089-2023 was initiated by Siemens Healthcare Diagnostics, Inc. on Aug 24, 2022 and is designated Class II. Reason for recall: Dimension Vista¿ High Density Lipoprotein Cholesterol (HDLC) Potential for Erroneous Result. If an HDLC QC or patient result is obtained without an Abnormal Reaction [E145]: flag, a falsely… The recall status is ongoing. Affected quantity: 1592 units.
Recall Details
- Product Type
- Devices
- Report Date
- Oct 26, 2022
- Initiation Date
- Aug 24, 2022
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 1592 units
Product Description
Dimension Vista High Density Lipoprotein Cholesterol (HDLC) Flex reagent cartridge, IVD Siemens Material Number (SMN): 10464340 Catalog Number: K3048A
Reason for Recall
Dimension Vista¿ High Density Lipoprotein Cholesterol (HDLC) Potential for Erroneous Result. If an HDLC QC or patient result is obtained without an Abnormal Reaction [E145]: flag, a falsely depressed or elevated result ranging from -90% to 133% may be observed.
Distribution Pattern
Nationwide Foreign: Canada
Code Information
UDI-DI : 0084276802043821250BA2022090710464340840 Lot Number: 21250BA Exp. Date: 2022-09-07