Z-0089-2023 Class II Ongoing

Recalled by Siemens Healthcare Diagnostics, Inc. — Newark, DE

Recall Details

Product Type
Devices
Report Date
October 26, 2022
Initiation Date
August 24, 2022
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1592 units

Product Description

Dimension Vista High Density Lipoprotein Cholesterol (HDLC) Flex reagent cartridge, IVD Siemens Material Number (SMN): 10464340 Catalog Number: K3048A

Reason for Recall

Dimension Vista¿ High Density Lipoprotein Cholesterol (HDLC) Potential for Erroneous Result. If an HDLC QC or patient result is obtained without an Abnormal Reaction [E145]: flag, a falsely depressed or elevated result ranging from -90% to 133% may be observed.

Distribution Pattern

Nationwide Foreign: Canada

Code Information

UDI-DI : 0084276802043821250BA2022090710464340840 Lot Number: 21250BA Exp. Date: 2022-09-07