Z-0093-2023 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- October 26, 2022
- Initiation Date
- August 30, 2022
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 373 PCK
Product Description
N Antiserum to Human IgG, 2mL Variant for BN II System and BN ProSpec System-IVD Siemens Material Number (SMN): 10446297
Reason for Recall
N Antiserum to Human IgG are not meeting the current High-Dose Hook Effect expectation of up to 648 mg/L for IgGU (urine) and up to 2290 mg/L for IgGC (CSF).
Distribution Pattern
Nationwide
Code Information
UDI-DI: 0405686900179VF Lot Number: 153088 Future lots may be impacted by the issue but will then be provided with a corresponding instruction note within the reagent package