Z-0093-2023 Class II Ongoing

Recalled by SIEMENS HEALTHCARE DIAGNOSTICS INC — Tarrytown, N/A

Recall Details

Product Type
Devices
Report Date
October 26, 2022
Initiation Date
August 30, 2022
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
373 PCK

Product Description

N Antiserum to Human IgG, 2mL Variant for BN II System and BN ProSpec System-IVD Siemens Material Number (SMN): 10446297

Reason for Recall

N Antiserum to Human IgG are not meeting the current High-Dose Hook Effect expectation of up to 648 mg/L for IgGU (urine) and up to 2290 mg/L for IgGC (CSF).

Distribution Pattern

Nationwide

Code Information

UDI-DI: 0405686900179VF Lot Number: 153088 Future lots may be impacted by the issue but will then be provided with a corresponding instruction note within the reagent package