Z-0093-2023 Class II Ongoing

Recalled by SIEMENS HEALTHCARE DIAGNOSTICS INC — Tarrytown, N/A

FDA device recall Z-0093-2023 was initiated by SIEMENS HEALTHCARE DIAGNOSTICS INC on Aug 30, 2022 and is designated Class II. Reason for recall: N Antiserum to Human IgG are not meeting the current High-Dose Hook Effect expectation of up to 648 mg/L for IgGU (urine) and up to 2290 mg/L for IgGC (CSF). The recall status is ongoing. Affected quantity: 373 PCK.

Recall Details

Product Type
Devices
Report Date
Oct 26, 2022
Initiation Date
Aug 30, 2022
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
373 PCK

Product Description

N Antiserum to Human IgG, 2mL Variant for BN II System and BN ProSpec System-IVD Siemens Material Number (SMN): 10446297

Reason for Recall

N Antiserum to Human IgG are not meeting the current High-Dose Hook Effect expectation of up to 648 mg/L for IgGU (urine) and up to 2290 mg/L for IgGC (CSF).

Distribution Pattern

Nationwide

Code Information

UDI-DI: 0405686900179VF Lot Number: 153088 Future lots may be impacted by the issue but will then be provided with a corresponding instruction note within the reagent package

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0093-2023
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.