Z-0094-2023 Class II Ongoing

Recalled by SIEMENS HEALTHCARE DIAGNOSTICS INC — Tarrytown, N/A

Recall Details

Product Type
Devices
Report Date
October 26, 2022
Initiation Date
August 30, 2022
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
10,201 packs

Product Description

N Antiserum to Human IgG, 5mL Variant for BN II System and BN ProSpec System- IVD Siemens Material Number (SMN): 10446299

Reason for Recall

N Antiserum to Human IgG are not meeting the current High-Dose Hook Effect expectation of up to 648 mg/L for IgGU (urine) and up to 2290 mg/L for IgGC (CSF).

Distribution Pattern

Nationwide

Code Information

UDI-DI: 0405686900179VF Lot Number: 153083; 153085; 153087A; 153090; 153095B; 153099A; 153004B. Future lots may be impacted by the issue but will then be provided with a corresponding instruction note within the reagent package