Z-0095-2023 Class II Ongoing

Recalled by Becton Dickinson & Company — Franklin Lakes, NJ

Recall Details

Product Type
Devices
Report Date
October 26, 2022
Initiation Date
September 15, 2022
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
2,196,000

Product Description

BD Vacutainer Tube Sodium Fluoride: 100 mg, Potassium Oxalate: 20 mg, Catalog Number 367001. For collecting, transporting, separating, and processing blood for blood alcohol determination.

Reason for Recall

The affected product contains isobutylene which has recently demonstrated potential interference in determination of methanol by gas chromatography methods. The interference may lead to false positive results in methanol testing.

Distribution Pattern

Distribution US nationwide, Canada, New Zealand, and Hong Kong.

Code Information

UDI-DI: Case: 50382903670012; Shelf: 30382903670018; Each: 00382903670017; ALL LOTS are affected.