Product Description
Atellica CH Lithium_2 (LITH_2), Siemens Material Number 11532401. in vitro diagnostic assay
Reason for Recall
Potential for falsely elevated LITH_2 and Li results on the Atellica CH 930 Analyzer. In the worst case, there is a potential for erroneously elevated lithium patient results, which may lead to inappropriate dosage adjustment, potentially affecting treatment efficacy.
Distribution Pattern
Worldwide distribution - US Nationwide.
Code Information
UDI-DI: 00630414287935 All lots impacted: Lots 110238 exp 10/18/22 110302 exp 12/13/22 120017 exp 4/18/23