Z-0104-2023 Class II Ongoing

Recalled by Siemens Healthcare Diagnostics, Inc. — Tarrytown, NY

Recall Details

Product Type
Devices
Report Date
October 26, 2022
Initiation Date
September 19, 2022
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1342 US; 2266 OUS

Product Description

Atellica CH Lithium_2 (LITH_2), Siemens Material Number 11532401. in vitro diagnostic assay

Reason for Recall

Potential for falsely elevated LITH_2 and Li results on the Atellica CH 930 Analyzer. In the worst case, there is a potential for erroneously elevated lithium patient results, which may lead to inappropriate dosage adjustment, potentially affecting treatment efficacy.

Distribution Pattern

Worldwide distribution - US Nationwide.

Code Information

UDI-DI: 00630414287935 All lots impacted: Lots 110238 exp 10/18/22 110302 exp 12/13/22 120017 exp 4/18/23