Z-0116-2013 Class II Terminated

Recalled by Advandx Inc — Woburn, MA

FDA device recall Z-0116-2013 was initiated by Advandx Inc on Sep 14, 2012 and is designated Class II. Reason for recall: Some strains of Pseudomonas putida cross-react with the P. aeruginosa PNA probe used in E. coli/P. aeruginosa PNA FISH (KT007) due to a sequence similarity resulting in a cross-reaction that may… The recall status is terminated (terminated Mar 5, 2013). Affected quantity: 33 units.

Recall details

Product Type
Devices
Report Date
Oct 31, 2012
Initiation Date
Sep 14, 2012
Termination Date
Mar 5, 2013
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
33 units

Product description

AdvanDx E. coli/P. aeruginosa PNA FISH, In Vitro Diagnostic Culture Identification Kit for Escherichia coli and Pseudomonas aeruginosa Catalog Number : KT007 Usage: E. coli/P. aeruginosa PNA FISH is a multicolor, qualitative nucleic acid hybridization assay intended for identification of Escherichia coli and Pseudomonas aeruginosa on smears of positive blood cultures containing Gram-negative rods. The E.coli/P. aeruginosa PNA FISH assay is indicated for use as an aid in the diagnosis of E. coli and/or P. aeruginosa bacteremia.

Reason for recall

Some strains of Pseudomonas putida cross-react with the P. aeruginosa PNA probe used in E. coli/P. aeruginosa PNA FISH (KT007) due to a sequence similarity resulting in a cross-reaction that may produce a red-positive signal

Distribution pattern

Worldwide distribution: USA (nationwide) including: CA, FL, IL, MI, and NJ; and country of: Great Britain.

Code information

Lot Numbers: 01812A-US, 03412A-US and 00512A-EU Exp Date: 12-2013

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0116-2013
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Advandx Inc? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-0116-2013. Report a data issue: hello@regdatalab.com.