Z-2138-2014 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- August 13, 2014
- Initiation Date
- July 4, 2014
- Termination Date
- September 30, 2014
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 10 kits
Product Description
AdvanDx Quick FISH Fixation Kit for BC ( Blood Cultures), 50 tests. In vitro diagnostic. The QuickFISH Fixation Kits are components that are sold separately for the Staphylococcus QuickFISH BC test Catalog number: QFFixBC1-50 (a component for QFSTABC1-50)
Reason for Recall
Positive result in the Internal negative control well for the Staphylococcus QuickFISH BC test, failing QC
Distribution Pattern
Distributed in the states of AZ, ID, FL, MO, OH, MI, NJ, and VA.
Code Information
Lot Number: 04914 Exp. Date: 2015-06