Z-0133-2023 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- November 2, 2022
- Initiation Date
- September 20, 2022
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 162
Product Description
da Vinci X (IS4200) and Xi (IS4000) systems, which consists of finished devices Surgeon Console (SSC), Patient Cart (PSCART), Vision System (VSS)
Reason for Recall
Inadvertent energy delivery from surgical system instrument if 1) Force bipolar and bipolar instruments installed on system 2) Force bipolar connected to Force Triad generator, bipolar not connected to generator 3) Yellow pedal associated with force bipolar is pressed, released with head in surgeon console 4) Blue pedal associated with bipolar is pressed, resulting in force bipolar energy delivery
Distribution Pattern
US: MS, PA, WI, KY, TN, MI, FL, GA, NC, NV, LA, OH, IL, MO, IN, CA, AK, MD, TX, IA, MA, NE, AZ, WA, CO, AL, SC
Code Information
Software version P10 (OS4 v10.0.0/A70_P10_B738) SOFTWARE,EMBEDDED RLS,IS4000 AND IS4200,A70_P10_B738 (Model: 610092-738) on Model Name/Model Number/UDI-DI: ASSY,PSCART,IS4000,4-ARM,P10/380652-55/00886874110720, ASSY,SSC,IS4000,P10/380677-28/00886874110744, ASSY,VSS VISION SYSTEM,IS4000,P10/381121-45/00886874110898, ASSY,PSCART,IS4200,P10/380620-47/00886874115404, ASSY,VSC,IS4000,RECONDITIONED,P10/380721-15/00886874115343, ASSY,SSC,IS4000,RECONDITIONED,P10/380723-16/00886874115374