Z-0134-2021 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- October 28, 2020
- Initiation Date
- August 25, 2020
- Termination Date
- May 4, 2023
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- US:5970 units ; OUS: 1446 units
Product Description
Siemens Dimension Vista Akaline Phosphatase (ALPI) Flex reagent cartridge - In vitro diagnostic test for the quantitative measurement of alkaline phosphatase in human serum and plasma Catalog Number: K2115, SMN# 10642444
Reason for Recall
Low outlier resultsgenerated with Alkaline Phosphatase (ALPI) Flex reagent cartridge.The maximum difference observed with a patient sample was approximately -27% at a concentration of 129 U/L (2.15 ukat/L). The low outlier results are generated from the last five (5) tests of the well
Distribution Pattern
Worldwide distribution - US Nationwide and the countries of Canada, Australia, Austria, Belgium, Bahamas, Czech Republic, Denmark, France, Germany, Italy, Japan, Korea, Netherlands, Norway, New Zealand, Portugal, Slovakia, Slovenia, Spain, Switzerland.
Code Information
Lot Number/UDI/Exp. Date: 1) 19247AB UDI: 0084276803903419247AB20090310642444840 Exp.2020-09-03, (2) 19282BB UDI: 0084276803903419282BB20100810642444840 Exp. 2020-10-08, (3) 19330BD UDI:0084276803903419330BD20112510642444840 Exp. 2020-11-25. Expanded Recall: 3/10/21 20337AB, 20311BA, 20275AA, 20218AA, 20203BA, 20170AA, 20125BA, 20078AB.