Z-0164-2027 Class II Ongoing

Recalled by Howmedica Osteonics Corp. — Mahwah, NJ

FDA device recall Z-0164-2027 was initiated by Howmedica Osteonics Corp. on Sep 16, 2026 and is designated Class II. Reason for recall: Reprocessing instructions provided in LSTPI-B specify a sterilization dry time of 30 minutes for the trays sterilized in blue wrap. When blue wrap is used, the 30-minute dry time may lead to a small… The recall status is ongoing. Affected quantity: 605 trays.

Recall Details

Product Type
Devices
Report Date
Oct 21, 2026
Initiation Date
Sep 16, 2026
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
605 trays

Product Description

TS Universal Baseplate Finishing Tray, Triathlon. Catalog Number: 6543-6-506.

Reason for Recall

Reprocessing instructions provided in LSTPI-B specify a sterilization dry time of 30 minutes for the trays sterilized in blue wrap. When blue wrap is used, the 30-minute dry time may lead to a small amount of residual moisture remaining after sterilization.

Distribution Pattern

These trays are provided by sale or loan. Loans are provided through Distributor. Foreign Consignees will be identified by the countries impacted. 158 trays have been distributed OUS.

Code Information

Catalog Number: 6543-6-506. GTIN: 07613327548082. All lot numbers used prior to the correction date of May 8, 2026 are impacted.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0164-2027
openFDA file date
Oct 26, 2026
Loaded by RegDataLab
Oct 27, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.